The sequence is characterized by a large spike of all four colors and is caused by either a micro-air bubble or small debris migrating out of the capillary into the laser path and therefore scattering the light. For example, the spike is at base 450 and is masking the true base at this location. The best solution is to run the sample again on the DNA Analyzer.